|
|
|
| How
to prepare IR samples? |
|
1.
Materials for sample preparation
The following table lists the typical
IR materials for various applications. These IR materials
are used for IR window or diluents. It is worthy noting
that hygroscopic materials are suitable for organic
samples while non-hygroscopic materials are used for
water-containing samples.
|
| |
2. Sample preparation for transmission
analysis
2. 1 Solid samples
(2.1.1) Thin plate samples
You can use standard magnetic plate sample holder.
|
|
(2.1.2) Powder samples
You can press powder into a pellet. In this method,
you can select different diluents (matrixes) for various
applications. For mid-IR frequency range, KBr, KCl or
diamond dust can be used. For far-infrared testing,
high-density polyethylene (HDPE) or diamond dust is
suitable. For near-infrared analysis, CsI or KBr can
be selected. You can prepare a KBr Pellet as follows:
- (a) The powder sample and KBr must
be ground to reduce the particle size to less than
5 mm in diameter. Otherwise, large particles scatter
the infrared beam and cause a slope baseline of spectrum.
(b) Add a spatula full of KBr into an agate mortar
and grind it to fine powder until crystallites can
no longer be seen and it becomes somewhat “pasty”
and sticks to the mortar.
|
|
- (c) Take a small amount of powder sample
(about of 0.1-2% of the KBr amount, or just enough
to cover the tip of spatula) mix with the KBr powder.
Subsequently grind the mixture for 3-5 minutes.
(d) Assemble the die-set as shown in the drawing.
When assembling the die, please add the powder to
the 7mm collar. Put the die together with the powder
into the Qwik Handi-Press. Press the powder for 2
minutes to form a pellet. A good KBr pellet is thin
and transparent. Opaque pellets give poor spectra,
because little infrared beam passes through them.
White spots in a pellet indicate that the powder is
not ground well enough, or is not dispersed properly
in the pellets.
(e) Disassemble the die set and take out the 7mm collar.
Put the collar together with the pellet onto the sample
holder.
(f) Please clean the die after you finish your experiment.
It is worthy noting that Br- from the
KBr can often replace ligands in the compound whose
spectrum is desired. |
2.2. Liquid samples
(2.2.1) Making a sandwich
To prepare a liquid sample to IR analysis, firstly place
a drop of the liquid on the face of a highly polished
salt plate (such as NaCl, AgCl or KBr), then place a second
plate on top of the first plate so as to spread the liquid
in a thin layer between the plates, and clamps the plates
together. Finally wipe off the liquid out of the edge
of plate. You can mount the sandwich plate onto the sample
holder. After finishing your experiment, clean the plates
with isopropanol (No water!) and return them to the desiccators. |
|
Precaution:
Volatile liquid can’t be prepared with this method,
because it will evaporate while its spectrum is being
obtained. If the liquid sample is toxic or smelly, please
don’t use this method. In addition, NaCl and KBr
are dissolved into water, and thus they can’t
be used for aqueous samples.
(2.2.2) Using a liquid cell
The thermo Nicolet demountable pathlength cell is designed
for liquid sample. Assemble the cell as shown in the
drawing. Then use a syringe to fill the liquid into
the cell. Finally seal the cell. |
|
3. Sample preparation
for ATR analysis
The strongly infrared adsorbing materials such as rubbers,
polymer films and aqueous solutions can be analyzed by
the attenuated total reflection (ATR) technique. Minimum
or no sample preparation is required for ATR analysis.
There are two kinds of crystals available at our laboratory,
i.e., 45º-ZnSe and 45º-Ge. 3.1
Solid samples
For solid samples, please use the flat crystal. To obtain
the best results, the recommended sample dimension is
75-90mm long ´ 16-25mm wide ´ <4mm thick.
The sample surface must be flat, otherwise intimate
contact between the sample and the ATR crystal can not
be obtained. |
|
Firstly place a sample on
top surface of the crystal. Then place the gripper plate
on the sample. Adjust the pressure applied to the gripper
plate to ensure the consistent contact is achieved between
the crystal and the sample. Precaution: if you find
a considerable force is necessary to overcome the resistance,
STOP! Excessive pressure can crack and permanent damage
the crystal. Also, the crystal must be keep clean and
scratch-free. Scratches affect the sample/crystal contact
and thus have significant influence on the depth of
penetration.
3.2 Liquid samples
For liquid samples, please use the trough crystal. Fill
the trough plate with your liquid sample. Then place
the cover over the trough plate.
Precaution:
Ensure the liquid sample does not attack the
crystal.
After you have finished your experiment,
clean up the crystal using proper solvents (Acetone
or water can be used for ZnSe and Ge).
4. Sample preparation for PM-IRRAS
measurement
Typically, you can deposit a metal film on the regular
glass slide (or silicon wafer). The metal/glass combination
can be acted as the reference sample. In addition, you
will deposit a monolayer (or organic film) over the
metal film. Hence, you will have two samples, i.e.,
the metal/glass combination and the monolayer/metal/glass
composite. The recommended dimension of sample is 55~75mm
x 17~25mm. |
References:
Sampling catalogue, Thermo Nicolet. |
| Related
subtopics: |
What
is FT-IR?
What is ATR?
What is Diffuse Reflectance Spectroscopy? What is PM-IRRAS?
How to prepare IR samples?
What does FT-IR do?
How to run FT-IR? |
|
|
© 2009 NU ANCE. All rights reserved.
Northwestern University, 2220 Campus Drive #2036, Evanston, IL 60208-3108
Phone: 847-467-2318, Fax: 847-467-6573
Email questions to: nuance@northwestern.edu. |
|
|