The sampling surface is
pressed into intimate optical contact with the top
surface of the crystal such as ZnSe or Ge. The IR
radiation from the spectrometer enters the crystal.
It then reflects through the crystal and penetrating
“into” the sample a finite amount with
each reflection along the top surface via the so-called
“evanescent” wave. At the output end of
the crystal, the beam is directed out of the crystal
and back into the normal beam path of the spectrometer.
To obtain internal
reflectance, the angle of incidence must exceed the
so-called ‘critical’ angle. This angle
is a function of the real parts of the refractive
indices of both the sample and the ATR crystal:

Where n2
is the refractive index of the sample and n1
is the refractive index of the crystal. The evanescent
wave decays into the sample exponentially with distance
from the surface of the crystal over a distance on
the order of microns. The depth of penetration of
the evanescent wave d is defined as the distance
form the crystal-sample interface where the intensity
of the evanescent decays to 1/e(37%) of its
original value. It can be given by:

Where l is the
wavelength of the IR radiation. For instance, if the
ZnSe crystal (n1=2.4) is used,
the penetration depth for a sample with the refractive
index of 1.5 at 1000cm-1 is estimated to
be 2.0µm when the angle of incidence is 45°.
If the Ge crystal (n1=4.0) is
used under the same condition, the penetration depth
is about 0.664µm. The depth of penetration and
the total number of reflections along the crystal
can be controlled either by varying the angle of incidence
or by selection of crystals. Different crystals have
different refractive index of the crystal material.
By the way, it is worthy noting that different crystals
are applied to different transmission range (ca.
ZnSe for 20,000~650cm-1, Ge for 5,500~800cm-1).