|
|
|
| What
is FT-IR? |
|
| Infrared (IR) spectroscopy
is a chemical analytical technique, which measures the
infrared intensity versus wavelength (wavenumber) of light.
Based upon the wavenumber, infrared light can be categorized
as far infrared (4 ~ 400cm-1), mid infrared
(400 ~ 4,000cm-1) and near infrared (4,000
~ 14,000cm-1). |
Molecular
vibration modes induced by IR beam
|
|
|
|
|
Infrared spectroscopy detects the vibration
characteristics of chemical functional groups in a
sample. When an infrared light interacts with the
matter, chemical bonds will stretch, contract and
bend. As a result, a chemical functional group tends
to adsorb infrared radiation in a specific wavenumber
range regardless of the structure of the rest of the
molecule. For example, the C=O stretch of a carbonyl
group appears at around 1700cm-1 in a variety
of molecules. Hence, the correlation of the band wavenumber
position with the chemical structure is used to identify
a functional group in a sample. The wavenember positions
where functional groups adsorb are consistent, despite
the effect of temperature, pressure, sampling, or
change in the molecule structure in other parts of
the molecules. Thus the presence of specific functional
groups can be monitored by these types of infrared
bands, which are called group wavenumbers.
The early-stage IR instrument
is of the dispersive type, which uses a prism or a
grating monochromator. The dispersive instrument is
characteristic of a slow scanning. A Fourier Transform
Infrared (FTIR) spectrometer obtains infrared spectra
by first collecting an interferogram of a sample signal
with an interferometer, which measures all of infrared
frequencies simultaneously. An FTIR spectrometer acquires
and digitizes the interferogram, performs the FT function,
and outputs the spectrum.
|
| |
Schematic
illustration of FTIR system
|
An interferometer utilizes
a beamsplitter to split the incoming infrared beam into
two optical beams. One beam reflects off of a flat mirror
which is fixed in place. Another beam reflects off of
a flat mirror which travels a very short distance (typically
a few millimeters) away from the beamsplitter. The two
beams reflect off of their respective mirrors and are
recombined when they meet together at the beamsplitter.
The re-combined signal results from the “interfering”
with each other. Consequently, the resulting signal
is called interferogram, which has every infrared frequency
“encoded” into it. When the interferogram
signal is transmitted through or reflected off of the
sample surface, the specific frequencies of energy are
adsorbed by the sample due to the excited vibration
of function groups in molecules. The infrared signal
after interaction with the sample is uniquely characteristic
of the sample. The beam finally arrives at the detector
and is measure by the detector. The detected interferogram
can not be directly interpreted. It has to be “decoded”
with a well-known mathematical technique in term of
Fourier Transformation. The computer can perform the
Fourier transformation calculation and present an infrared
spectrum, which plots adsorbance (or transmittance)
versus wavenumber.
When an interferogram is Fourier transformed,
a single beam spectrum is generated. A single beam spectrum
is a plot of raw detector response versus wavenumber.
A single beam spectrum obtained without a sample is
called a background spectrum, which is induced by the
instrument and the environments. Characteristic bands
around 3500 cm-1 and 1630 cm-1
are ascribed to atmospheric water vapor, and the bands
at 2350 cm-1 and 667 cm-1 are
attributed to carbon dioxide. A background spectrum
must always be run when analyzing samples by FTIR. When
an interferogram is measured with a sample and Fourier
transformed, a sample single beam spectrum is obtained.
It looks similar to the background spectrum except that
the sample peaks are superimposed upon the instrumental
and atmospheric contributions to the spectrum. To eliminate
these contributions, the sample single beam spectrum
must be normalized against the background spectrum.
Consequently, a transmittance spectrum is obtained as
follows.
%T = I/Io
Where %T is transmittance; I
is the intensity measured with a sample in the beam
(from the sample single beam spectrum); Io
is the intensity measured from the back ground spectrum
The absorbance spectrum can be calculated
from the transmittance spectrum using the following
equation.
A = -log10 T
Where A is the absorbance.
The final transmittance/absorbance spectrum
should be devoid of all instrumental and environmental
contributions, and only present the features of the
sample. If the concentrations of gases such as water
vapor and carbon dioxide in the instrument are the same
when the background and sample spectra are obtained,
their contributions to the spectrum will ratio out exactly
and their bands will not occur. If the concentrations
of these gases are different when the background and
sample spectra are obtained, their bands will appear
in the sample spectrum. |
Reference:
Brian C. Smith, Fundamentals of Fourier Transform Infrared
spectroscopy, CRC press, Boca Raton, 1996. |
| Related
subtopics: |
What
is FT-IR?
What is ATR?
What is Diffuse Reflectance
Spectroscopy? What is PM-IRRAS?
How to prepare IR samples?
What does FT-IR do?
How to run FT-IR? |
|
|
© 2008 NU ANCE. All rights reserved.
Northwestern University, 2220 Campus Drive #2036, Evanston, IL 60208-3108
Phone: 847-467-2318, Fax: 847-467-6573
Email questions to: nuance@northwestern.edu. |
|
|